Ms Sinergy offers the service in this variable for the calibration of digital manometers, analog manometers, pressure transmitters, and liquid columns with a measurement range from 0 inH2O to 4 inH2O.
| Item | Code | Magnitude | Headquarters | Measurement Range | Instrument to be calibrated | Uncertainty | Instruments/Standards | Normative Document | Method | Personnel | Duration |
|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | DG8 | Pressure | Carrera 7 Este 1A-32 Interior 1 | 0 Pa ≤ p ≤ 996 Pa, 0 inH2O ≤ p ≤ 4 inH2O | Differential pressure manometer (digital - analog - transmitters) accuracy class ≥ ± 0.5% of full scale | 1,0 Pa (0,0041 inH2O) | Digital differential pressure manometers, Accuracy class ± 0.5% of full scale | Guideline DKD R-6-1 Calibration of Pressure Measuring Instruments Edition 03/2014 Revision 3. Except for Section 8.5 | Direct comparison | Jhon Alexander Ortiz Preciado, José Luis Rodriguez Roa | 35.00 |
| 2 | DG8 | Pressure | In situ | 0 Pa ≤ p ≤ 996 Pa, 0 inH2O ≤ p ≤ 4 inH2O | Differential pressure manometer (digital - analog - transmitters) accuracy class ≥ ± 0.5% of full scale | 1,0 Pa (0,0041 inH2O) | Digital differential pressure manometers, Accuracy class ± 0.5% of full scale | Guideline DKD R-6-1 Calibration of Pressure Measuring Instruments Edition 03/2014 Revision 3. Except for Section 8.5 | Direct comparison | Jhon Alexander Ortiz Preciado, José Luis Rodriguez Roa | 35.00 |
| 3 | DG8 | Pressure | Carrera 7 Este 1A-32 Interior 1 | 0 Pa ≤ p ≤ 50 Pa, 0 inH2O ≤ p ≤ 0,2 inH2O | Liquid column manometers accuracy class ≥ ± 3% of full scale | 1,1 Pa (0,0045 inH2O) | Digital differential pressure manometers, Accuracy class ± 0.5% of full scale | Procedure ME-021 for the calibration of manometric and barometric liquid columns Spanish Metrology Center Digital Edition 2. 2020 | Direct comparison | Jhon Alexander Ortiz Preciado, José Luis Rodriguez Roa | 50.00 |
| 4 | DG8 | Pressure | Carrera 7 Este 1A-32 Interior 2 | 50 Pa < p ≤ 124 Pa, 0,2 inH2O < p ≤ 0,5 inH2O | Liquid column manometers accuracy class ≥ ± 3% of full scale | 1,8 Pa (0,0072 inH2O) | Digital differential pressure manometers, Accuracy class ± 0.5% of full scale | Procedure ME-021 for the calibration of manometric and barometric liquid columns Spanish Metrology Center Digital Edition 2. 2020 | Direct comparison | Jhon Alexander Ortiz Preciado, José Luis Rodriguez Roa | 50.00 |
| 5 | DG8 | Pressure | Carrera 7 Este 1A-32 Interior 3 | 124 Pa < p ≤ 747 Pa, 0,5 inH2O < p ≤ 3 inH2O | Liquid column manometers accuracy class ≥ ± 3% of full scale | 3,0 Pa (0,012 inH2O) | Digital differential pressure manometers, Accuracy class ± 0.5% of full scale | Procedure ME-021 for the calibration of manometric and barometric liquid columns Spanish Metrology Center Digital Edition 2. 2020 | Direct comparison | Jhon Alexander Ortiz Preciado, José Luis Rodriguez Roa | 50.00 |
| 6 | DG8 | Pressure | In situ | 0 Pa ≤ p ≤ 50 Pa, 0 inH2O ≤ p ≤ 0,2 inH2O | Liquid column manometers accuracy class ≥ ± 3% of full scale | 1,1 Pa (0,0046 inH2O) | Digital differential pressure manometers, Accuracy class ± 0.5% of full scale | Procedure ME-021 for the calibration of manometric and barometric liquid columns Spanish Metrology Center Digital Edition 2. 2020 | Direct comparison | Jhon Alexander Ortiz Preciado, José Luis Rodriguez Roa | 50.00 |
| 7 | DG8 | Pressure | In situ | 50 Pa < p ≤ 124 Pa, 0,2 inH2O < p ≤ 0,5 inH2O | Liquid column manometers accuracy class ≥ ± 3% of full scale | 1,8 Pa (0,0073 inH2O) | Digital differential pressure manometers, Accuracy class ± 0.5% of full scale | Procedure ME-021 for the calibration of manometric and barometric liquid columns Spanish Metrology Center Digital Edition 2. 2020 | Direct comparison | Jhon Alexander Ortiz Preciado, José Luis Rodriguez Roa | 50.00 |
| 8 | DG8 | Pressure | In situ | 124 Pa < p ≤ 747 Pa, 0,5 inH2O < p ≤ 3 inH2O | Liquid column manometers accuracy class ≥ ± 3% of full scale | 3,0 Pa (0,012 inH2O) | Digital differential pressure manometers, Accuracy class ± 0.5% of full scale | Procedure ME-021 for the calibration of manometric and barometric liquid columns Spanish Metrology Center Digital Edition 2. 2020 | Direct comparison | Jhon Alexander Ortiz Preciado, José Luis Rodriguez Roa | 50.00 |